• DocumentCode
    1226011
  • Title

    A novel method for assessing position-sensitive detector performance

  • Author

    Clinthorne, N.H. ; Rogers, W.L. ; Shao, L. ; Hero, A.O., III. ; Koral, K.F.

  • Author_Institution
    Michigan Univ., Ann Arbor, MI, USA
  • Volume
    36
  • Issue
    1
  • fYear
    1989
  • Firstpage
    1127
  • Lastpage
    1131
  • Abstract
    A marked-point process model of a position-sensitive detector is developed which includes the effects of detector efficiency, spatial response, energy response, and source statistics. The average mutual information between the incident distribution of gamma -rays and the detector response is derived and used as a performance index of detector optimization. A brief example is presented which uses this figure-of-merit for optimization of light-guide dimensions for the SPRINT IT modular scintillation camera.<>
  • Keywords
    biomedical equipment; radioisotope scanning and imaging; SPRINT IT modular scintillation camera; detector efficiency; detector optimization; energy response; gamma -rays; light-guide dimensions; marked-point process model; mutual information; position-sensitive detector; source statistics; spatial response; Cameras; Distortion measurement; Entropy; Mutual information; Position sensitive particle detectors; Probability distribution; Random variables; Relays; Uncertainty; Upper bound;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.34617
  • Filename
    34617