DocumentCode :
1226018
Title :
Design and imaging studies of a position sensitive photomultiplier based dynamic SPECT system
Author :
Singh, Manbir ; Leahy, Richard ; Brechner, Ricardo ; Yan, Xiao-hong
Author_Institution :
Dept. of Radiol., Univ. of Southern California, Los Angeles, CA, USA
Volume :
36
Issue :
1
fYear :
1989
fDate :
2/1/1989 12:00:00 AM
Firstpage :
1132
Lastpage :
1137
Abstract :
The design and image reconstruction aspects of dynamic SPECT (single-photon-emission computed tomography) instrumentation utilizing multiple rings of position-sensitive photomultiplier tubes (PSPMTs) are reported. A diverging cone-beam collimator is attached to each PSPMT to achieve a 22-cm-diameter spherical field of view in a system designed for head imaging. The resolution with the diverging collimator is estimated to be approximately 20% worse than with a parallel hold collimator. Iterative and direct cone-beam reconstruction algorithms have been investigated under different conditions of angular sampling and counting statistics. The ramifications of temporal resolution and signal-to-noise ratio are discussed in relation to dynamic SPECT imaging of labeled cisplatin for chemotherapy
Keywords :
biomedical equipment; computerised picture processing; computerised tomography; photomultipliers; radioisotope scanning and imaging; chemotherapy; cone-beam reconstruction algorithms; diverging cone-beam collimator; dynamic SPECT; head imaging; image reconstruction aspects; labeled cisplatin; parallel hold collimator; position-sensitive photomultiplier tubes; signal-to-noise ratio; single-photon-emission computed tomography; temporal resolution; Collimators; Computed tomography; Head; Image reconstruction; Instruments; Photomultipliers; Reconstruction algorithms; Sampling methods; Signal resolution; Statistics;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/23.34618
Filename :
34618
Link To Document :
بازگشت