Title :
Interstitial antennas tipped with reactive load
Author :
Ahn, Hee-Ran ; Lee, Kwyro
Author_Institution :
Dept. of Electr. Eng. & Comput. Sci., Korea Adv. Inst. of Sci. & Technol., Taejon, South Korea
Abstract :
New and simple interstitial antennas tipped with reactive load (RL) are presented. For their design, a matching technique based on transmission line theory is also suggested. To verify the technique, one immersed in muscle phantom is designed, fabricated, measured, and compared. The measured reflection coefficient at 2.45 GHz is -28.4 dB, which is considered the best among those reported, and the measured region greater than 43°C is a rugby ball (major axis 4.5cm and minor axis 2.3cm), which demonstrates that it can be applied for a deep seated tumor or cancer.
Keywords :
UHF antennas; biomedical equipment; cancer; hyperthermia; microwave heating; muscle; phantoms; radiation therapy; transmission line theory; tumours; 2.45 GHz; SAR distributions; cancer; hyperthermia; interstitial antennas; muscle phantom; reactive load; rugby ball; specific absorption rate; transmission line equivalent circuit; transmission line theory; tumor; Antenna measurements; Cancer; Equivalent circuits; Imaging phantoms; Loaded antennas; Muscles; Neoplasms; Transmission line antennas; Transmission line measurements; Transmission line theory;
Journal_Title :
Microwave and Wireless Components Letters, IEEE
DOI :
10.1109/LMWC.2004.842826