Curvature-based nonfiducial registration for the frameless stereotactic operating microscope
Author
Friets, Eric M. ; Strohbehn, John W. ; Roberts, David W.
Author_Institution
Thayer Sch. of Eng., Dartmouth Coll., Hanover, NH, USA
Volume
42
Issue
9
fYear
1995
Firstpage
867
Lastpage
878
Abstract
The frameless stereotactic operating microscope permits information from CT and MRI scans to be displayed in the operating microscope in the proper scale and perspective without using a mechanical frame in the process, when the microscope is positioned over the surgical field. This registration is currently done using fiducials that must be present when the imaging scans are taken. A new technique, based on the theory of curvatures, has been developed as an alternative to the use of fiducials for the registration of a patient´s head position during surgery with diagnostic images from CT or MRI. Surface curvatures are estimated from a trace of points of the surface of the skin, obtained from both the diagnostic images and an intraoperative nonimaging ultrasonic rangefinder. The point traces need not be identical or evenly spaced. Plots of the resulting curvature fields are compared visually and the alignment determined. Phantom testing, using a human skull, has resulted in a median alignment error of 1.95 mm. Testing using a human subject with clinically obtained data has resulted in alignment errors on the order of 7 mm.
Keywords
biomedical NMR; biomedical equipment; computerised tomography; image registration; medical image processing; surgery; CT scans; MRI scans; alignment error; curvature-based nonfiducial registration; diagnostic images; frameless stereotactic operating microscope; human skull; mechanical frame; patient´s head position; phantom testing; point traces; surface curvatures; Computed tomography; Head; Humans; Imaging phantoms; Magnetic resonance imaging; Microscopy; Skin; Skull; Surgery; Testing; Computers; Equipment Design; Humans; Magnetic Resonance Imaging; Mathematics; Microscopy; Phantoms, Imaging; Skull; Stereotaxic Techniques; Surgical Equipment; Tomography, X-Ray Computed;
fLanguage
English
Journal_Title
Biomedical Engineering, IEEE Transactions on
Publisher
ieee
ISSN
0018-9294
Type
jour
DOI
10.1109/10.412654
Filename
412654
Link To Document