DocumentCode
1229682
Title
An AC Josephson Voltage Standard for AC–DC Transfer-Standard Measurements
Author
Merimaa, M. ; Nyholm, K. ; Vainio, Markku ; Lassila, Antti
Author_Institution
Centre for Metrol. & Accreditation, Espoo
Volume
56
Issue
2
fYear
2007
fDate
4/1/2007 12:00:00 AM
Firstpage
239
Lastpage
243
Abstract
We describe the traceability chain of length measurements at Centre for Metrology and Accreditation (MIKES) from atomic clocks to the lasers of primary interferometers. Crucial part of the traceability chain, an optical frequency comb generator linking optical frequencies to atomic clocks, is described in detail. The frequency comb generator is used in frequency calibrations of iodine-stabilized lasers, which are operated in compliance with the recommendations of the practical realization of the definition of the meter. Measured absolute frequencies of iodine-stabilized lasers, time records of the measurements, and the respective Allan deviations demonstrate the solid performance of the MIKES laser frequency standards. The results are in good agreement with the recommended values, as well as with the independent characterizations of the measured lasers
Keywords
calibration; frequency standards; laser frequency stability; laser variables measurement; measurement by laser beam; Allan deviations; MIKES; atomic clocks; iodine-stabilized lasers; laser frequency calibrations; laser frequency standards; length measurements; optical frequencies; optical frequency comb generator; primary interferometers; Atom optics; Atomic clocks; Frequency measurement; Length measurement; Measurement standards; Metrology; Optical interferometry; Solid lasers; Time measurement; Voltage; AC measurements; Josephson arrays; Josephson devices; ac voltage standard;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.2007.891153