DocumentCode :
1229745
Title :
Rise-Time Calibration of 50-GHz Sampling Oscilloscopes: Intercomparison Between PTB and NPL
Author :
Bieler, Mark ; Seitz, Steffen ; Spitzer, Meinhard ; Hein, Günter ; Pierz, Klaus ; Siegner, Uwe ; Basu, Anjan M. ; Smith, Andrew J A ; Harper, Matthew R.
Author_Institution :
Phys.-Technische Bundesanstalt, Braunschweig
Volume :
56
Issue :
2
fYear :
2007
fDate :
4/1/2007 12:00:00 AM
Firstpage :
266
Lastpage :
270
Abstract :
The paper presents the results of the first intercomparison of the rise-time calibration of a 50-GHz sampling oscilloscope between Physikalisch-Technische Bundesanstalt, Germany, and National Physical Laboratory, U.K. (EUROMET project 641). Rise-times between 6.0 and 6.6 ps were obtained using optoelectronic techniques. Detailed uncertainty budgets are given, which yield typical rise-time uncertainties of 1.2 ps. The absolute values of the degree of equivalence are les 0.3 ps with an expanded uncertainty of 1.0 ps and, thus, validate the experimental techniques
Keywords :
calibration; oscilloscopes; 50 GHz; 6.0 to 6.6 ps; EUROMET project 641; Germany; National Physical Laboratory; Physikalisch-Technische Bundesanstalt; U.K.; electro-optic sampling; optoelectronic techniques; rise-time calibration; sampling oscilloscopes; Bandwidth; Calibration; Earth Observing System; Laboratories; Optical pulse generation; Oscilloscopes; Pulse measurements; Sampling methods; Switches; Voltage; 50 GHz; Calibration; electrooptic sampling (EOS); intercomparison; oscilloscope; rise-time;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.2007.890609