• DocumentCode
    1230268
  • Title

    Atomic Vector Gradiometer System Using Cesium Vapor Cells for Magnetocardiography: Perspective on Practical Application

  • Author

    Kim, Kiwoong ; Lee, Won-Kyu ; Kim, In-Seon ; Moon, Han Seb

  • Author_Institution
    Bio-signal Res. Lab., Korea Res. Inst. of Stand. & Sci., Daejeon
  • Volume
    56
  • Issue
    2
  • fYear
    2007
  • fDate
    4/1/2007 12:00:00 AM
  • Firstpage
    458
  • Lastpage
    462
  • Abstract
    We developed an optical pumping magnetometer system for magnetocardiography (MCG). The basic system requirements for measuring MCG are as follows: a 5-muT weak bias field in a magnetically shielded room, a 20-mm cell for spatial resolution, and a 30-Hz bandwidth for heart signals. In order to reduce the environmental magnetic noise, a double-cell gradiometer configuration has been adopted. The measured noise level was ~10 pT/radicHz. By switching current flows through a symmetrical three-axis rectangular Helmholtz coil set, the system is designed to measure three vector components of the magnetic field gradient. We show that this configuration reduces the error in current dipole localization with respect to the usual normal component detection by computer simulation of confidence-region estimation. Based on the simulation, we suggest that the system is applicable for diagnosing myocardial infarction on the anterior wall of a human heart in coronary artery disease
  • Keywords
    biomedical measurement; caesium; magnetic field measurement; magnetocardiography; magnetometers; 20 mm; 30 Hz; 5 muT; anterior wall; atomic vector gradiometer system; cesium vapor cells; coronary artery disease; dipole localization; double-cell gradiometer configuration; environmental magnetic noise; heart signals; human heart; magnetic field gradient; magnetocardiography; myocardial infarction; normal component detection; optical pumping magnetometer system; rectangular Helmholtz coil set; Atomic measurements; Bandwidth; Heart; Magnetic field measurement; Magnetic noise; Magnetic shielding; Magnetometers; Noise measurement; Optical pumping; Spatial resolution; Atomic magnetometer; magnetocardiography (MCG); myocardial infarction; optical pumping magnetometer;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.2007.890610
  • Filename
    4126884