Title : 
Computer-Controlled Inductive Bridge Standards for Calibrating AC Resistance Thermometry Bridges at Low Frequencies
         
        
            Author : 
Ramm, Günther ; Moser, Harald ; Vollmer, Reinhold
         
        
        
        
        
            fDate : 
4/1/2007 12:00:00 AM
         
        
        
        
            Abstract : 
 This paper describes the setup and calibration of inductive bridge standards which realize selected voltage ratios—in particular, ratios between 0.1 and 1.3—at frequencies of a few tens of hertz with uncertainties from   to
  to   . With these bridge standards, ac thermometry bridges can be calibrated with the lowest possible uncertainties either manually or using computer control.
 . With these bridge standards, ac thermometry bridges can be calibrated with the lowest possible uncertainties either manually or using computer control.  
         
        
            Keywords : 
Bridge circuits; Calibration; Electrical resistance measurement; Frequency; Resistors; Standards development; Temperature; Thermal resistance; Uncertainty; Voltage; AC thermometry bridge (TB); bridge standard; inductive voltage divider; international temperature scale;
         
        
        
            Journal_Title : 
Instrumentation and Measurement, IEEE Transactions on
         
        
        
        
        
            DOI : 
10.1109/TIM.2007.891068