Title :
Accuracy Improvement of Real-Time Load-Pull Measurements
Author :
Teppati, Valeria ; Ferrero, Andrea ; Parena, Daniela ; Pisani, Umberto
Author_Institution :
Dipt. di Elettronica, Politecnico di Torino
fDate :
4/1/2007 12:00:00 AM
Abstract :
This paper describes a new procedure aimed to improve the effectiveness of real-time load-pull calibration. Load-pull measurement accuracy is strongly affected by calibration residual uncertainty. The novel methodology reduces this uncertainty contribution by means of error terms optimization. The proposed method has been tested with simulations and applied to actual measurement data. Considerable improvements have been achieved
Keywords :
calibration; measurement uncertainty; microwave devices; microwave measurement; network analysers; accuracy improvement; calibration residual uncertainty; directional couplers; error terms optimization; microwave devices; microwave measurements; microwave phase shifters; microwave power amplifiers; real-time load-pull calibration; real-time load-pull measurements; tuners; Calibration; Impedance measurement; Microwave devices; Optimization methods; Performance evaluation; Power measurement; Real time systems; Reflection; Testing; Tuners; Directional couplers; microwave devices; microwave measurements; microwave phase shifters; microwave power amplifiers; tuners; uncertainty;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
DOI :
10.1109/TIM.2007.891049