DocumentCode
1233057
Title
Analysis of second-order harmonic distortion of ADC using bispectrum
Author
Gao, Xiao Ming ; Ovaska, Seppo J. ; Shenghe, Sun ; Jenq, Y.C.
Author_Institution
Dept. of Inf. Technol., Lappeenranta Univ. of Technol., Finland
Volume
45
Issue
1
fYear
1996
fDate
2/1/1996 12:00:00 AM
Firstpage
50
Lastpage
55
Abstract
This paper presents a new frequency-domain approach to test the second-order harmonic distortion of an analog-to-digital converter (ADC). It uses triple correlation and its Fourier transform, bispectrum, to suppress quantization noise and improve the testing accuracy of weak second-order harmonics. Our approach is independent of the triggering delay that is a known weakness of the spectral averaging method. The new method is also less sensitive to the quantization noise than the power spectrum averaging method. Both theoretical analyses and illustrative computer simulations are provided
Keywords
analogue-digital conversion; digital simulation; discrete Fourier transforms; frequency-domain analysis; harmonic analysis; harmonic distortion; ADC; Fourier transform; analog-to-digital converter; bispectrum; computer simulation; frequency-domain analysis; quantization noise; second-order harmonic distortion; second-order harmonics; spectral averaging; testing accuracy; triggering delay; triple correlation; Discrete Fourier transforms; Distortion measurement; Fourier transforms; Harmonic analysis; Harmonic distortion; Power harmonic filters; Quantization; Signal to noise ratio; Spectral analysis; Testing;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/19.481311
Filename
481311
Link To Document