Influence of field inhomogeneity in magnetic calibration coils
Author :
Nissen, Johan ; Paulsson, Lars-Erik
Author_Institution :
Swedish Radiat. Protection Inst., Stockholm, Sweden
Volume :
45
Issue :
1
fYear :
1996
fDate :
2/1/1996 12:00:00 AM
Firstpage :
304
Lastpage :
306
Abstract :
Calibration of magnetic field detectors in the Helmholtz coil can be performed with an absolute accuracy of some tenths of a percent. This paper presents a correction, verified experimentally, that in some cases can be of the same order of magnitude. The correction, which is due to the inhomogeneity of the magnetic flux density from the Helmholtz coil over the detector coil to be calibrated, is found to exceed 1 percent if the size of the Helmholtz coil is twice the size of the detector coil
Keywords :
calibration; coils; magnetic flux; magnetic sensors; Helmholtz coil; correction; detector coil; field inhomogeneity; inhomogeneity; magnetic calibration coils; magnetic field detectors; magnetic flux density; Calibration; Coils; Detectors; Laboratories; Magnetic field measurement; Magnetic fields; Magnetic flux; Magnetic flux density; Magnetometers; Protection;
fLanguage :
English