Abstract
An unbalanced transformer bridge designed for inductance standard calibration is described. Basic theory of the bridge is discussed in which two components of the unbalanced voltage are measured and used to determine a ratio of compared inductance standards. Significant sources of errors are analyzed and numerical limits given. The uncertainty of comparison, in the frequency range 100 Hz-2 kHz, is assessed to be smaller than 30 ppm for 10 mH-1 H standards and smaller than 100 ppm for 1 mH-10 mH standards. The bridge is convenient for automatic operation
Keywords
bridge circuits; bridge instruments; calibration; inductance measurement; measurement errors; measurement standards; 100 Hz to 2 kHz; automatic operation; errors; inductance standard calibration; numerical limits; unbalanced transformer bridge; unbalanced voltage; uncertainty; Bridge circuits; Calibration; Detectors; Frequency; Inductance measurement; Measurement standards; Particle measurements; Phase detection; Phase measurement; Voltage;