• DocumentCode
    123972
  • Title

    Verification of Robotic Surgery Tasks by Reachability Analysis: A Comparison of Tools

  • Author

    Bresolin, Davide ; Geretti, Luca ; Muradore, Riccardo ; Fiorini, Paolo ; Villa, Tania

  • Author_Institution
    Univ. of Bologna, Bologna, Italy
  • fYear
    2014
  • fDate
    27-29 Aug. 2014
  • Firstpage
    659
  • Lastpage
    662
  • Abstract
    In this paper we discuss the application of formal methods for the verification of properties of control systems designed for autonomous robotic systems. We illustrate our proposal in the context of surgery by considering the automatic execution of a simple action such as puncturing. To prove that a sequence of subtasks planned on pre-operative data can successfully accomplish the surgical operation despite model uncertainties, we specify the problem by using hybrid automata. We express the requirements of interest as questions about reachability properties of the hybrid automaton model. Then, we compare the different performance of current state-of-the art tools for reachability analysis of hybrid automata.
  • Keywords
    automata theory; control system synthesis; medical robotics; reachability analysis; surgery; control system design; formal methods; hybrid automata; puncturing action; reachability analysis; robotic surgery tasks; Accuracy; Approximation methods; Automata; End effectors; Safety; Surgery; Formal Verification; Hybrid Systems; Robotic Surgery;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Digital System Design (DSD), 2014 17th Euromicro Conference on
  • Conference_Location
    Verona
  • Type

    conf

  • DOI
    10.1109/DSD.2014.55
  • Filename
    6927307