Abstract
Calibration is a crucial step in structured-light-based range sensing device. The step involves the determination of the intrinsic parameters of both the camera and the projector that constitute the device and the extrinsic parameters between the two instruments. The traditional solution requires the use of an external calibration object with an accurately measured pattern printed on it. This paper presents a calibration design that makes use of a liquid-crystal display (LCD) panel as the calibration object. The LCD panel´s planarity is of industrial grade and is thus dependable. The pattern shown on the LCD panel is programmable and is thus convenient to produce in high precision. We show that, with the design, the projector-and-camera system parameters can be calibrated with far fewer images with much higher accuracy. Extensive experiments are shown to illustrate the dramatic improvement in performance.
Keywords
calibration; cameras; liquid crystal displays; optical projectors; LCD panel; calibration; liquid-crystal display panel; projector-camera system; structured-light-based range sensing system; Calibration; liquid-crystal display (LCD) panel; projector-and-camera system; structured light system;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.2008.925016
Filename
4538270
Link To Document