• DocumentCode
    1242715
  • Title

    An inverted-microstrip resonator for human head proton MR imaging at 7 tesla

  • Author

    Zhang, Xiaoliang ; Ugurbil, Kamil ; Sainati, Robert ; Chen, Wei

  • Author_Institution
    Dept. of Radiol., Univ. of Minnesota, Minneapolis, MN, USA
  • Volume
    52
  • Issue
    3
  • fYear
    2005
  • fDate
    3/1/2005 12:00:00 AM
  • Firstpage
    495
  • Lastpage
    504
  • Abstract
    As an extension of the previously developed microstrip transmission line (MTL) RF coil design, a high-frequency RF volume coil using multiple inverted MTL (iMTL) resonators for human head imaging at high magnetic field strength of 7 tesla (T) is reported. Compared to conventional MTL resonators, iMTL resonators can operate at higher frequency with lower losses and, thus, are suitable for designs of high-frequency RF volume coils with large coil size for human MR imaging and spectroscopy at high fields. An approach using capacitive terminations was analyzed and applied to the design of the iMTL volume coil for improving RF field homogeneity and broadening frequency-tuning range. A performance-comparison study was conducted between the prototype iMTL volume coil and a custom-built TEM volume coil at 7 T. The iMTL volume coil presents a comparable SNR and intrinsic B1 homogeneity to the TEM volume coil. Phantom and the human head images acquired using the iMTL volume coil are also presented. The proposed iMTL volume coil provides an efficient and alternative solution to design high-frequency and large-size volume coils for human MR applications at very high fields.
  • Keywords
    biomedical MRI; microstrip resonators; phantoms; 7 T; capacitive terminations; custom-built TEM volume coil; human head proton MR imaging; inverted microstrip transmission line volume coil; inverted-microstrip resonator; phantom; Coils; Humans; Magnetic fields; Magnetic heads; Microstrip resonators; Protons; Prototypes; Radio frequency; Spectroscopy; Transmission lines; High field; MR imaging; MTL volume coil; RF coil; human brain; inverted microstrip transmission line resonator; Brain; Equipment Design; Equipment Failure Analysis; Head; Humans; Magnetic Resonance Imaging; Magnetic Resonance Spectroscopy; Magnetics; Phantoms, Imaging; Protons; Reproducibility of Results; Sensitivity and Specificity; Transducers;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.2004.842968
  • Filename
    1396390