DocumentCode
1244134
Title
Long-term hermeticity and biological performance of anodically bonded glass-silicon implantable packages
Author
Harpster, Timothy J. ; Nikles, Stefan A. ; Dokmeci, Mehmet R. ; Najafi, Khalil
Author_Institution
Center for Wireless Integrated Microsystems, Univ. of Michigan, Ann Arbor, MI, USA
Volume
5
Issue
3
fYear
2005
Firstpage
458
Lastpage
466
Abstract
This paper reviews long-term test results obtained from a series of tests on glass-silicon (Si) hermetically sealed packages. Results are presented from 1) a 9.9-year ongoing room temperature phosphate-buffered saline (PBS) soak test of four packages; 2) accelerated soak tests in high temperature saline of 28 samples resulting in an extrapolated mean-time-to-failure (MTTF) at 37°C of 177 years; 3) a 2.7-year in vitro 97°C PBS soak test of a single package; and 4) in situ hermeticity and biocompatibility tests from 12 packages implanted in four guinea pigs-three packages implanted in two guinea pigs (each) for 1 month and another two guinea pigs for 20 and 22 months. All of the packages remained hermetically sealed over the lifetime of the implant. A detailed histological report of the implants is provided suggesting that they elicit no profound adverse reaction from the body.
Keywords
electronics packaging; hermetic seals; prosthetics; accelerated soak tests; biocompatibility tests; biocompatible package; glass-silicon implantable packages; hermetic package; in vivo testing; microsystems packages; phosphate buffered saline soak test; Bonding; Glass; Hermetic seals; Implants; Integrated circuit packaging; Rough surfaces; Surface roughness; Temperature; Testing; Thermal management; Biocompatible package; hermetic package; in vivo testing; microsystems packages;
fLanguage
English
Journal_Title
Device and Materials Reliability, IEEE Transactions on
Publisher
ieee
ISSN
1530-4388
Type
jour
DOI
10.1109/TDMR.2005.854374
Filename
1545908
Link To Document