Abstract :
The electrical properties of a radiotransparent carbon electrode were evaluated and compared with those of a high-quality Ag/AgCl electrode. In order to approach clinical electrocardiographic (ECG) recording conditions, special experimental arrangements for the determination of impedance, phase angle, and noise characteristics were used. It is shown that the carbon electrode provides precordial ECG signals of excellent quality during catheterizations and allows simultaneous use of X-ray examination.
Keywords :
biomedical equipment; carbon; electrocardiography; electrodes; laboratory apparatus and techniques; Ag-AgCl electrode; ECG recordings; catheterization laboratory; electrical properties; impedance; noise characteristics; phase angle; precordial ECG signals; radiotransparent C electrode; Biomedical electrodes; Cardiology; Catheterization; Electrocardiography; Heart; Laboratories; Manufacturing; Plastics; Rubber; Wire; Carbon; Electric Conductivity; Electrocardiography; Electrodes; Heart Catheterization; Humans; Silver;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/10.1402
Filename :
1402
Link To Document :
بازگشت