Title :
Electrical impedance tomography: induced-current imaging achieved with a multiple coil system
Author :
Gençer, Nevzat G. ; Ider, Y. Ziya ; Williamson, Samuel J.
Author_Institution :
Dept. of Phys., New York Univ., NY, USA
Abstract :
An experimental study of induced-current electrical impedance tomography verifies that image quality is enhanced by employing six rather than three induction coils by increasing the number of independent measurements. However, with an increasing number of coils, the inverse problem becomes more sensitive to measurement noise. Using 16 electrodes to measure surface voltages, it is possible to collect 6×15=90 independent measurements. For comparison purposes, images of two-dimensional conductivity perturbations are reconstructed by using the data for three and six coils with the truncated pseudoinverse algorithm. By searching for the optimal truncation index that minimizes the noise error plus the resolution error, the signal-to-noise ratio of the data acquisition system was established as 58 db. Images obtained with this six-coil system reveal the sizes and locations of the conductivity perturbations. This system also provides images within the central region of the object space, a capability not achieved in previous experimental studies using only three circular coils. Nevertheless, the three-coil system can identify the conductivity perturbations near the periphery. However, it displays shifts in the locations and spread in the sizes of perturbations near the center of the object.
Keywords :
biomedical equipment; coils; electric impedance imaging; inverse problems; 58 dB; data acquisition system; electrical impedance tomography; image quality enhancement; induced-current imaging; measurement noise; medical diagnostic imaging; multiple coil system; noise error; optimal truncation index; resolution error; signal-to-noise ratio; surface voltages measurement; truncated pseudoinverse algorithm; two-dimensional conductivity perturbations; Coils; Conductivity; Electric variables measurement; Image quality; Impedance measurement; Inverse problems; Noise measurement; Signal to noise ratio; Surface impedance; Tomography; Algorithms; Analog-Digital Conversion; Artifacts; Data Display; Electric Conductivity; Electric Impedance; Electrodes; Equipment Design; Humans; Image Enhancement; Image Processing, Computer-Assisted; Models, Theoretical; Signal Processing, Computer-Assisted; Tomography;
Journal_Title :
Biomedical Engineering, IEEE Transactions on