• DocumentCode
    1247616
  • Title

    Coded pulse excitation for ultrasonic strain imaging

  • Author

    Liu, Jie ; Insana, Michael F.

  • Author_Institution
    Dept. of Biomed. Eng., California Univ., Davis, CA, USA
  • Volume
    52
  • Issue
    2
  • fYear
    2005
  • Firstpage
    231
  • Lastpage
    240
  • Abstract
    Decorrelation strain noise can be significantly reduced in low echo-signal-to-noise (eSNR) conditions using coded excitation. Large time-bandwidth-product (>30) pulses are transmitted into tissue mimicking phantoms with 2.5-mm diameter inclusions that mimic the elastic properties of breast lesions. We observed a 5-10 dB improvement in eSNR that led to a doubling of the depth of focus for strain images with no reduction of spatial resolution. In high eSNR conditions, coded excitation permits the use of higher carrier frequencies and shorter correlation windows to improve the attainable spatial resolution for strain relative to that obtained with conventional short pulses. This paper summarizes comparative studies of strain imaging in noise-limited conditions obtained by short pulses and four common aperiodic codes (chirp, Barker, suboptimal, and Golay) as a function of attenuation, eSNR and applied strain. Imaging performance is quantified using SNR for displacement (SNR/sub d/), local modulation transfer function (LMTF), and contrast-to-noise ratio for strain (CNR/sub /spl epsi//). We found that chirp and Golay codes are the most robust for imaging soft tissue deformation using matched filter decoding. Their superior performance is obtained by balancing the need for low-range lobes, large eSNR improvement, and short-code duration.
  • Keywords
    Golay codes; biomechanics; biomedical ultrasonics; decoding; noise; ultrasonic imaging; 2.5 mm; breast lesions; carrier frequencies; coded pulse excitation; contrast-noise ratio; decorrelation strain noise; elastic properties; filter decoding; imaging performance; local modulation transfer function; spatial resolution; strain images; tissue deformation; tissue mimicking phantoms; ultrasonic strain imaging; Breast; Capacitive sensors; Chirp; Decorrelation; Focusing; Imaging phantoms; Lesions; Noise reduction; Spatial resolution; Ultrasonic imaging; Animals; Humans; Mathematics; Neoplasms; Phantoms, Imaging; Sensitivity and Specificity; Signal Processing, Computer-Assisted; Transducers; Ultrasonography; Ultrasonography, Mammary;
  • fLanguage
    English
  • Journal_Title
    Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0885-3010
  • Type

    jour

  • DOI
    10.1109/TUFFC.2005.1406549
  • Filename
    1406549