    PET Demonstrator for a Human Brain Scanner Based on Monolithic Detector Blocks

  • Author

    Sarasola, I. ; Mendes, P. Rato ; De Acilu, P. Garéca ; Cañadas, M. ; Vela, O. ; Cela, J.M. ; Oller, J.C. ; Nunez, L. ; Willmott, C. ; Pérez, J.M.

  • Author_Institution
    Centro de Investig. Energeticas, Medioambientales y Tecnol. (CIEMAT), Madrid, Spain
  • Volume
    58
  • Issue
    5
  • fYear
    2011
  • Firstpage
    2190
  • Lastpage
    2197
  • Abstract
    We have implemented and evaluated a positron emission tomography (PET) demonstrator using two monolithic detector blocks operating in coincidence with dedicated application-specific integrated circuit (ASIC) readout. Each detector is composed of a monolithic lutetium yttrium orthosilicate (LYSO) scintillator coupled to a pair of Hamamatsu S8550-02 APD arrays. The front-end electronics of this demonstrator is based on the VATA240 ASIC readout, which sums the charge provided by each row and column of the APD array. The ASIC has been characterized obtaining a noise per row or column less than 2000 electrons rms with the APD at its inputs and a good linear response in the range from 5 fC to 30 fC. We have acquired energy spectra of 22Na and 137Cs radioactive sources, achieving energy resolutions between 13.2% and 14.1% full width at half maximum (FWHM) at 511 keV. We have estimated the interaction position over the surface of the monolithic blocks using Neural Networks (NN) position determining algorithms, obtaining spatial resolutions at the detector level down to 2.1 mm FWHM. By using this detector technology and electronics we have achieved images of point sources with spatial resolutions as good as 2.1 mm FWHM for filtered back projection (FBP) reconstructions methods with single slice rebinning (SSRB). Based on the results obtained with this demonstrator, we are developing a PET insert for existing magnetic resonance imaging (MRI) equipment, to be installed in a collaborating hospital and used for clinical PET-MRI of the human brain.
  • Keywords
    application specific integrated circuits; biomedical MRI; brain; nuclear electronics; positron emission tomography; solid scintillation detectors; 137Cs radioactive sources; 22Na radioactive sources; Hamamatsu S8550-02 APD arrays; LYSO scintillator; PET demonstrator; PET insert; SSRB; VATA240 ASIC readout; application-specific integrated circuit readout; clinical PET-MRI; coincidence operation; energy spectra; filtered back projection reconstruction methods; front-end electronics; human brain scanner; magnetic resonance imaging equipment; monolithic blocks; monolithic detector blocks; monolithic lutetium yttrium orthosilicate; neural networks position determining algorithms; point sources; positron emission tomography; single slice rebinning; Application specific integrated circuits; Crystals; Detectors; Laboratories; Magnetic resonance imaging; Positron emission tomography; Spatial resolution; Application specific integrated circuit; artificial neural networks; position sensitive detectors; positron emission tomography;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2011.2152856
  • Filename
    5898438