DocumentCode :
1252597
Title :
Measurement error for six port ANA
Author :
Bergeault, E. ; Huyart, B. ; Geneves, G. ; Jallet, L.
Author_Institution :
Lab. Central des Ind. Electr., Fontenay Aux Roses, France
Volume :
137
Issue :
6
fYear :
1990
fDate :
12/1/1990 12:00:00 AM
Firstpage :
415
Lastpage :
416
Abstract :
Although six-port ANA theory requires that the input impedances of power detectors remain constant with incident power, this is not generally the case if diode detectors are used. Consequently, uncertainties in the measurement of the unknown load reflection coefficient appear. Measurements of the variations of diode impedances are used in a computer simulation to estimate these uncertainties.
Keywords :
calibration; computerised instrumentation; electric impedance; measurement errors; microwave reflectometry; network analysers; 2 GHz; automatic network analyser; computer simulation; diode detectors; diode impedances; load reflection coefficient; measurement error; power detectors; six port ANA; six port reflectometer;
fLanguage :
English
Journal_Title :
Microwaves, Antennas and Propagation, IEE Proceedings H
Publisher :
iet
ISSN :
0950-107X
Type :
jour
Filename :
62508
Link To Document :
بازگشت