DocumentCode :
1252627
Title :
Evaluation of a voltage source with three calculable RMS outputs
Author :
Lapuh, R. ; Svetik, Z.
Author_Institution :
Inst. of Quality & Metrol., Ljubljana, Slovenia
Volume :
46
Issue :
4
fYear :
1997
Firstpage :
784
Lastpage :
788
Abstract :
This paper describes a voltage source which generates three different waveforms. Measurement techniques with such a source are proposed, with emphasis on the uncertainty of each waveform used, when the ac amplitude is defined with a dc calibration of the source. The instrument circuitry and operation is described. Measurement results are also discussed. Emphasis is given to selected parameters of waveforms and to their effect on the measured rms value.
Keywords :
calibration; computerised instrumentation; digital-analogue conversion; measurement errors; microcomputer applications; signal sampling; waveform generators; RMS outputs; ac amplitude; calibration; dc calibration; direct digital sampling; rms value; root mean square; thermal convertor; uncertainty; voltage source; waveforms; AC generators; Calibration; Circuits; DC generators; Frequency; Instruments; Measurement techniques; Root mean square; Sampling methods; Voltage;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/19.650773
Filename :
650773
Link To Document :
بازگشت