A new structure transcutaneous transformer for artificial heart system
Author
Zhao, L. ; Foo, C.F. ; Tseng, K.J.
Author_Institution
Sch. of Electr. & Electron. Eng., Nanyang Technol. Inst., Singapore
Volume
35
Issue
5
fYear
1999
fDate
9/1/1999 12:00:00 AM
Firstpage
3550
Lastpage
3552
Abstract
A new geometrical design of a transcutaneous transformer for an artificial heart system is presented in this paper. The primary and secondary cores are made of ferrite and amorphous magnetic materials respectively. The experimental results indicate that it is possible to transmit an electrical power of 20 W which is sufficient to drive the heart pump. With a gap of 5 mm between the coils, its transmission efficiency with the inverter can be up to 90% with a switching frequency 102.5 kHz. Some characteristics of the proposed transformer are also shown and discussed in this paper
Keywords
coils; high-frequency transformers; pacemakers; transformer cores; 102.5 kHz; 20 W; 5 mm; amorphous magnetic material core; artificial heart system; ferrite core; full-bridge resonant convertor; geometrical design; heart pump drive; implanted coil; inverter; primary core; secondary core; transcutaneous transformer; Amorphous magnetic materials; Amorphous materials; Artificial heart; Circuit faults; Coils; Ferrites; Magnetic cores; Magnetostriction; Power supplies; Transformer cores;
fLanguage
English
Journal_Title
Magnetics, IEEE Transactions on
Publisher
ieee
ISSN
0018-9464
Type
jour
DOI
10.1109/20.800586
Filename
800586
Link To Document