• DocumentCode
    1260066
  • Title

    High precision measurement of DC voltage ratios from 20 V/10 V to 1000 V/10 V

  • Author

    Zhang, Xiuzeng ; Qie, Jiaping ; Zhang, Lili ; Shao, Haiming

  • Author_Institution
    Electromagn. Div., Nat. Inst. of Metrol., Beijing, China
  • Volume
    51
  • Issue
    1
  • fYear
    2002
  • fDate
    2/1/2002 12:00:00 AM
  • Firstpage
    59
  • Lastpage
    62
  • Abstract
    This paper presents a new method of calibrating de voltage ratios from 20 V: 10 V to 1000 V: 10 V. A new battery-powered differential voltage detector has been developed to reduce the uncertainty of the measurement system. An absolute self-calibration process was used and traceability to a voltage ratio standard is not necessary. The uncertainties of dc voltage ratios (20 V to 1000 V): 10 V were less than 2 × 10-7
  • Keywords
    calibration; stability; voltage dividers; voltage measurement; 1000 V to 10 V; 1000 to 10 V; 20 V to 10 V; 20 to 10 V; DC voltage divider; DC voltage ratio; absolute self-calibration; battery-powered differential voltage detector; reference divider; traceability; uncertainties; uncertainty; voltage ratio standard; Calibration; Capacitors; Circuits; Detectors; Preamplifiers; Resistors; Sampling methods; Stability; Temperature; Voltage measurement;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.989900
  • Filename
    989900