    An approach for measuring ultrasonic backscattering from biological tissues with focused transducers

  • Author

    Wang, Shyh-Hau ; Shung, K. Kirk

  • Author_Institution
    Bioeng. Program, Pennsylvania State Univ., University Park, PA, USA
  • Volume
    44
  • Issue
    7
  • fYear
    1997
  • fDate
    7/1/1997 12:00:00 AM
  • Firstpage
    549
  • Lastpage
    554
  • Abstract
    When the standard substitution method is used with a focused transducer to measure the backscattering coefficient from biological tissues including blood, it yields erroneous results. Extending the backscattering measurements to frequencies beyond 15 MHz necessitates the use of focused transducers because of the worsened signal-to-noise ratio-caused by the increased attenuation and the smaller transducer aperture size-in order to make the measurements close to the transducer. An approach which allows the use of focused transducers in backscattering measurements has been developed. It has been used to measure the backscattering coefficient of red cell suspensions of hematocrit ranging from a few percent to 30% in the frequency range from 5 MHz to 30 MHz. The results at hematocrits below 20% agree well with those obtained with the standard substitution method, although they differ as the hematocrit is increased beyond 20%. The experimental results also show that the fourth-power dependence of backscatter on frequency is in general approximately valid for suspended erythrocytes of hematocrit between 6% and 30%.
  • Keywords
    backscatter; bioacoustics; biological techniques; blood; cellular biophysics; ultrasonic scattering; 5 to 30 MHz; backscattering coefficient; biological tissues; focused transducer; fourth-power dependence; hematocrit; red cell suspensions; signal-to-noise ratio; suspended erythrocytes; ultrasonic backscattering measuring approach; Attenuation measurement; Backscatter; Biological tissues; Blood; Frequency measurement; Measurement standards; Signal to noise ratio; Size measurement; Ultrasonic transducers; Ultrasonic variables measurement; Animals; Erythrocytes; Hematocrit; Linear Models; Swine; Transducers; Ultrasonography, Doppler;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/10.594895
  • Filename
    594895