• DocumentCode
    1272595
  • Title

    A multiple-layer finite-element model of the surface EMG signal

  • Author

    Lowery, Madeleine M. ; Stoykov, Nikolay S. ; Taflove, Allen ; Kuiken, Todd A.

  • Author_Institution
    Rehabilitation Inst. of Chicago, Northwestern Univ., Chicago, IL, USA
  • Volume
    49
  • Issue
    5
  • fYear
    2002
  • fDate
    5/1/2002 12:00:00 AM
  • Firstpage
    446
  • Lastpage
    454
  • Abstract
    The effect of skin, muscle, fat, and bone tissue on simulated surface electromyographic (EMG) signals was examined using a finite-element model. The amplitude and frequency content of the surface potential were observed to increase when the outer layer of a homogeneous muscle model was replaced with highly resistive skin or fat tissue. The rate at which the surface potential decreased as the fiber was moved deeper within the muscle also increased. Similarly, the rate at which the surface potential decayed around the surface of the model, for a constant fiber depth, increased. When layers of subcutaneous fat of increasing thickness were then added to the model, EMG amplitude, frequency content, and the rate of decay of the surface EMG signal around the limb decreased, due to the increased distance between the electrodes and the active fiber. The influence of bone on the surface potential was observed to vary considerably, depending on its location. When located close to the surface of the volume conductor, the surface EMG signal between the bone and the source and directly over the bone increased, accompanied by a slight decrease on the side of the bone distal to the active fiber. The results emphasize the importance of distinguishing between the effects of material properties and the distance between source and electrode when considering the influence of subcutaneous tissue, and suggest possible distortions in the surface EMG signal in regions where a bone is located close to the skin surface.
  • Keywords
    biological tissues; biomedical electrodes; bone; electromyography; finite element analysis; physiological models; skin; surface potential; EMG amplitude; active fiber; bone tissue; constant fiber depth; electrodes; fat; fat tissue; frequency content; highly resistive skin; homogeneous muscle model; limb; material properties; multiple-layer finite-element model; muscle; outer layer; simulated surface electromyographic signals; skin; source; subcutaneous fat; subcutaneous tissue; surface EMG signal; surface potential; volume conductor; Bone tissue; Conducting materials; Distortion; Electrodes; Electromyography; Finite element methods; Frequency; Material properties; Muscles; Skin; Action Potentials; Adipose Tissue; Anisotropy; Bone and Bones; Computer Simulation; Electric Conductivity; Electromyography; Finite Element Analysis; Humans; Models, Neurological; Muscle, Skeletal; Reproducibility of Results; Skin;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/10.995683
  • Filename
    995683