Performance Measurements of the MicroPET FOCUS 120 for Iodine-124 Imaging
Author
Taleb, Dounia ; Bahri, Mohamed Ali ; Warnock, Geoffrey ; Salmon, Eric ; Luxen, André ; Plenevaux, Alain ; Anizan, Nadège ; Seret, Alain
Author_Institution
Exp. Med. Imaging Lab., Univ. of Liege, Liege, Belgium
Volume
59
Issue
5
fYear
2012
Firstpage
1868
Lastpage
1878
Abstract
This study aimed to evaluate the performance of the microPET FOCUS 120 for 124I in terms of counting rate capability and image quality using the NEMA NU 4-2008 methodology. Scanner sensitivity was measured for 124I for comparison and reached 75 cps/kBq, respectively, with the usual 350-650 keV energy window (EW) and 6 ns time window (TW). The noise equivalent count rate (NECR) index was defined as: NECR = RT2 /(RP +RGP) ( T = true, P = prompt, GP = γ-prompt). A rat phantom maximum NECR of 48 kcps was obtained for the 250-590 keV EW with 6 ns TW. An almost identical maximum NECR of 43 kcps was recorded for 350-590 and 350-650 keV EW and 6 ns TW. The 2 ns TW reduced the sensitivity and NECR by 40-50% for all EW. The mouse phantom NECR study was limited because of the maximum available activity concentration of 124I. The 250-590 keV EW showed the largest scatter and γ-prompt plus scatter fractions with 25.7% and 43%, respectively, for the rat phantom and 12.2% and 27% for the mouse phantom. With the 350-590 keV EW, these fractions decreased to 20% and 33.5% for the rat phantom and to 10% and 21% for the mouse phantom. The image quality was investigated with the NEMA NU 4-2008 dedicated phantom for four (two analytic and two iterative) 2D or 3D reconstruction methods. The lowest spillover ratios (SOR) for the phantom non-emitting regions were obtained for the 350-590 and 350-650 keV EWs. Recovery coefficients (RC) of the hot rods were the highest for the 350-590 keV EW except for the 1 mm rod. Scatter correction led to a large decrease in RC. The combination of the 350-590 keV EW with 6 ns TW appeared to be a good compromise between counting rate capability and image quality for the FOCUS 120, especially when maximum a posteriori reconstruction was used without scatter correction. Moreover this combination enabled the best quantification with an error as low as 0.36%.
Keywords
image denoising; image reconstruction; iodine; iterative methods; medical image processing; phantoms; positron emission tomography; sensitivity; 2D reconstruction methods; 3D reconstruction methods; I; NEMA NU 4-2008 methodology; counting rate capability; electron volt energy 250 keV to 650 keV; hot rods; image quality; iodine-124 imaging; iterative method; maximum a posteriori reconstruction; maximum available activity concentration; microPET FOCUS 120 performance measurements; mouse phantom NECR; noise equivalent count rate index; phantom nonemitting regions; rat phantom maximum NECR; recovery coefficients; scanner sensitivity; spillover ratios; time 2 ns; time 6 ns; Image quality; Image reconstruction; Isotopes; Mice; Phantoms; Sensitivity; Spatial resolution; Iodine-124; microPET; performance tests;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.2012.2205402
Filename
6304946
Link To Document