• DocumentCode
    1287857
  • Title

    Pinhole aperture design for 131I tumor imaging

  • Author

    Smith, Mark F. ; Jaszczak, Ronald J. ; Wang, Huili

  • Author_Institution
    Dept. of Biomed. Eng., Duke Univ., Durham, NC, USA
  • Volume
    44
  • Issue
    3
  • fYear
    1997
  • fDate
    6/1/1997 12:00:00 AM
  • Firstpage
    1154
  • Lastpage
    1160
  • Abstract
    A new pinhole aperture design for single photon imaging is introduced whose distinctive feature is a cylindrical channeled aperture. This allows greater attenuation near the pinhole aperture, which is especially important when imaging 131I with its medium and high energy emissions. The new design reduces to a conventional knife-edge aperture when the channel height is zero. Point source response functions were simulated for pinhole inserts with different pinhole diameters (0.4-2.0 mm diameter) and channel heights (0.1-3.0 mm) with acquisition parameters appropriate for 131I brain tumor imaging. A photon transport simulation code was used to comprehensively model photon transport and detection. Tradeoff curves between resolution and sensitivity suggest that varying pinhole diameter and channel height to achieve higher resolution at the expense of decreased sensitivity may be advantageous for imaging tumors with intratumoral administration of 131I radiolabeled monoclonal antibodies
  • Keywords
    biomedical equipment; brain; image resolution; iodine; medical image processing; photon transport theory; single photon emission computed tomography; 131I brain tumor imaging; 131I radiolabeled monoclonal antibodies; 131I tumor imaging; I; SPECT; acquisition parameters; attenuation; channel height; channel heights; conventional knife-edge aperture; cylindrical channeled aperture; high energy emission; intratumoral administration; medium energy emission; photon detection; photon transport simulation code; pinhole aperture design; pinhole diameters; pinhole inserts; point source response functions; resolution; sensitivity; single photon imaging; tradeoff curves; Apertures; Attenuation; Cancer; Collimators; Energy resolution; High-resolution imaging; Image resolution; Neoplasms; Optical imaging; Single photon emission computed tomography;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.596980
  • Filename
    596980