Title :
Robust heart sound activity detection in noisy environments
Author :
Sabarimalai Manikandan, M. ; Soman, K.P.
Author_Institution :
Center for Comput. Eng. & Networking, Amrita Univ., India
Abstract :
A novel and robust method for heart sound activity detection (HSAD) is presented. In this method, a new discriminative feature is developed based on the lag-1 autocorrelation coefficient and feature smoothing is introduced to identify the endpoints of low-level heart murmurs. Several experiments on a large scale phonocardiogram database show that the method significantly outperforms the other HSAD methods under varying levels of heart sounds and different types of noise. For a signal-to-noise ratio of 5 dB, the method achieves a sensitivity of 99.56% and a positive predictivity of 99.41% and has a smallest average time error value of 4.26 ms.
Keywords :
acoustic variables measurement; cardiology; feature extraction; medical signal detection; HSAD; average time error; feature smoothing; heart sound activity detection; lag-1 autocorrelation coefficient; phonocardiogram database; positive predictivity; signal-to-noise ratio; time 4.26 ms;
Journal_Title :
Electronics Letters
DOI :
10.1049/el.2010.1201