• DocumentCode
    1288061
  • Title

    The effect of radionuclide scatter in emission-transmission CT

  • Author

    Heanue, J.A. ; Brown, J.K. ; Tang, H.R. ; Hasegawa, B.H.

  • Author_Institution
    California Univ., San Francisco, CA, USA
  • Volume
    44
  • Issue
    3
  • fYear
    1997
  • fDate
    6/1/1997 12:00:00 AM
  • Firstpage
    1317
  • Lastpage
    1322
  • Abstract
    In order to assess the relative effect of scatter in comparison to attenuation and collimator blur, the authors imaged concentric cylinder, cold-lesion, and thorax-like phantoms with the emission transmission computed tomography (ETCT) system, a novel device capable of both X-ray CT and SPECT. For these experiments, the ETCT system was outfitted with a single channel of low-noise electronics which permitted acquisition of SPECT data with 3 keV FWHM energy resolution for good scatter rejection. Reference images were reconstructed with object-specific attenuation correction and accurate collimator response compensation These images were then compared to images which were reconstructed without attenuation correction, without collimator response, or with data that had been blurred to simulate 13 keV FWHM energy resolution. The results show that although scatter represents a measurable error, its effect is small compared to the effect of attenuation and collimator blur. As part of this study, the authors have also compared scatter rejection to the well-known dual energy window scatter correction method and to a technique which incorporates an estimate of the scatter distribution directly in the back-projector of the ML-EM algorithm. It is shown that scatter correction may be preferable to scatter rejection due to noise considerations
  • Keywords
    computerised tomography; single photon emission computed tomography; 13 keV; 3 keV; FWHM energy resolution; ML-EM algorithm; SPECT data; X-ray CT; attenuation blur; cold-lesion phantom; collimator blur; concentric cylinder phantom; emission-transmission CT; low-noise electronics; medical diagnostic imaging; noise considerations; radionuclide scatter effect; reference images; scatter distribution; scatter rejection; thorax-like phantom; Attenuation measurement; Collimators; Computed tomography; Electromagnetic scattering; Energy resolution; Image reconstruction; Particle scattering; Single photon emission computed tomography; X-ray imaging; X-ray scattering;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.597007
  • Filename
    597007