Developmental study on thermoluminescence dosimetry: two-step calibration instead of detector sensitivity factors
Author :
Toivonen, Matti J.
Author_Institution :
Finnish Centre for Radiat. & Nucl. Safety, Helsinki, Finland
Volume :
38
Issue :
3
fYear :
1991
fDate :
6/1/1991 12:00:00 AM
Firstpage :
901
Lastpage :
905
Abstract :
The author shows how to facilitate the use of solid thermoluminescence detectors (TLDs) in clinical measurement applications at therapy dose level by considering the two measurement cycles (i.e., the actual measurement cycle and the cycle intended for detector sensitivity normalization purposes) as a single process including self-correction for the errors caused by drift in reader sensitivity. The TLD system calibration, additional error sources of the integrated TLD process, a description of the experiments, and discussion of the results are presented
Keywords :
calibration; error correction; thermoluminescent dosimeters; detector sensitivity factors; error sources; self-correction; thermoluminescence detectors; thermoluminescence dosimetry; two-step calibration; Calibration; Detectors; Diodes; Dosimetry; Error correction; Hospitals; Laboratories; Medical treatment; Modems; Solids;