Title : 
A new self-balancing DC-substitution RF power meter
         
        
        
            Author_Institution : 
Electromagnetics Division, National Bureau of Standards, Boulder, CO 80302
         
        
        
        
        
        
            Abstract : 
Problems intrinsic in self-balancing Wheatstone bridges have led to the development of a new dc substitution microwave power meter. The new instrument allows four-terminal measurement of bolometer resistance and affords improved accuracy and lower noise at a lower cost than earlier instruments. Measurement errors due to imperfect behavior of the servo system are typically less than 0.01 percent.
         
        
            Keywords : 
Bolometers; Bridge circuits; Power measurement; Radio frequency; Resistance; Resistors; Voltmeters;
         
        
        
            Journal_Title : 
Instrumentation and Measurement, IEEE Transactions on
         
        
        
        
        
            DOI : 