DocumentCode :
1294035
Title :
A step-up method for calibrating inductive voltage dividers up to 1 MHz
Author :
Grohmann, Klaus
Author_Institution :
Physikalisch-Technische Bundesanstalt, Institut Berlin, 1000 Berlin 10, Germany
Issue :
4
fYear :
1976
Firstpage :
516
Lastpage :
518
Abstract :
A measuring method is presented which allows the calibration of inductive voltage dividers up to frequencies of 1 MHz. The characteristic element is an auxiliary divider with a 0.5-voltage ratio with adjustable foot and head tap voltages. Error sources due to incomplete guarding are discussed. The calibration uncertainties are ±0.3 ppm at 100 kHz and ±30 ppm at 1 MHz.
Keywords :
Calibration; Capacitance; Current measurement; Frequency measurement; Measurement uncertainty; Uncertainty; Voltage measurement;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.1976.6312278
Filename :
6312278
Link To Document :
بازگشت