DocumentCode
1294216
Title
The calibration and use of directional couplers without standards
Author
Allred, C.M. ; Manney, C.H.
Author_Institution
Electromagnetics Div., Nat. Bur. of Stand., Boulder, CO, USA
Issue
1
fYear
1976
fDate
3/1/1976 12:00:00 AM
Firstpage
84
Lastpage
89
Abstract
Basic principles and suggested techniques are presented for calibrating nonideal four-port couplers without standards. Under the assumption that the complex ratio of the side-arm voltages are known, all the parameters, except one of the directional couplers, can be determined without recourse to known values of impedances, shorts, opens, etc. Also, the surprising number of measurements that can be made with such an incompletely calibrated coupler are discussed.
Keywords
calibration; transmission lines; waveguide couplers; calibration; calibration technique; complex ratio; directional couplers; microwave couplers; nonideal four port couplers; parameter measurement; side arm voltages; Admittance; Admittance measurement; Directional couplers; Equations; Standards; Transmission line measurements;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.1976.6312311
Filename
6312311
Link To Document