    Absolute calibration of microwave attenuation measurement systems

  • Author

    Clark, R.F.

  • Author_Institution
    Phys. Div., Nat. Res. Council of Canada, Ottawa, Ont., Canada
  • Issue
    2
  • fYear
    1976
  • fDate
    6/1/1976 12:00:00 AM
  • Firstpage
    126
  • Lastpage
    128
  • Abstract
    The application of the Laverick 2:1 voltage step to the absolute calibration of microwave attenuation measurement systems is described. A practical example of a 10 GHz 6.0206-dB switched attenuator illustrates the contention that a very good attenuation measurement system can be checked within its own laboratory to an accuracy one or two orders of magnitude better than is usually possible in an intercomparison between laboratories.
  • Keywords
    attenuation measurement; calibration; microwave measurement; Laverick 2:1 voltage step; absolute calibration; microwave attenuation measurement systems; switched attenuator; Attenuation; Attenuation measurement; Attenuators; Microwave measurements; Microwave theory and techniques; Phase shifters; Receivers;
  • fLanguage
    English