• DocumentCode
    1295593
  • Title

    Lead and tungsten pinhole inserts for I-131 SPECT tumor imaging: experimental measurements and photon transport simulations

  • Author

    Smith, Mark F. ; Jaszczak, Ronald J. ; Wang, Huili ; Li, Jianying

  • Author_Institution
    Dept. of Biomed. Eng., Duke Univ., Durham, NC, USA
  • Volume
    44
  • Issue
    1
  • fYear
    1997
  • fDate
    2/1/1997 12:00:00 AM
  • Firstpage
    74
  • Lastpage
    82
  • Abstract
    The potential use of lead and tungsten pinhole inserts for high-resolution SPECT imaging of intratumor activity in I-131 radioimmunotherapy was investigated using experimental point source measurements and photon transport simulations. I-131 imaging is challenging because the primary photon emission is at 364 keV and penetration through the insert near the pinhole aperture is significant. Point source response functions (PSRF´s) for lead (Pb) and tungsten (W) pinhole inserts were measured experimentally. These response functions were simulated using a photon transport computer code that modeled the primary emission at 364 keV and secondary emissions at 284, 637, and 723 keV. Scatter within the pinhole insert, camera shielding, and scintillation crystal was modeled. There was good agreement between the experimental and simulated PSRF´s. Simulated point source response functions for geometrically identical Pb and W pinhole inserts were narrower for the W insert due to reduced penetration. SPECT pinhole imaging with these inserts was simulated for 3-cm-diameter tumors with a central core and 3-5-mm-thick shells. For one set of simulations there was no core activity, and for a second set the shell:core activity concentration ratio was 5:1. In both cases, the tumor shells were better resolved with the W insect. As a result, shell:core activity ratios were more accurate and contrast was improved with the use of the W pinhole insert. This study suggests that W inserts have potential advantages over Pb inserts for high-resolution I-131 pinhole imaging
  • Keywords
    iodine; lead; photon transport theory; radiation therapy; single photon emission computed tomography; tungsten; 284 to 723 keV; 3 cm; 3 to 5 mm; 3-cm-diameter tumors; I; I-131 SPECT tumor imaging; I-131 radioimmunotherapy; Pb; W; diagnostic nuclear medicine; high-resolution SPECT imaging; intratumor activity; lead pinhole inserts; photon transport computer code; photon transport simulations; point source response functions; primary emission; secondary emissions; therapeutic nuclear medicine; tungsten pinhole inserts; Apertures; Computational modeling; Computer simulation; Electromagnetic scattering; High-resolution imaging; Lead; Neoplasms; Optical computing; Single photon emission computed tomography; Tungsten;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.554828
  • Filename
    554828