• DocumentCode
    1296108
  • Title

    Generalized injected voltage method for the establishment of multidecade inductive voltage divider standard

  • Author

    Zhang, G.-M. ; Zhao, F.-Z.

  • Author_Institution
    Nat. Inst. of Metrol. of China. Beijing, China
  • Issue
    1
  • fYear
    1987
  • fDate
    3/1/1987 12:00:00 AM
  • Firstpage
    9
  • Lastpage
    12
  • Abstract
    A generalized injected voltage (GIV) method for a multidecade inductive voltage divider (IVD) is described. This method can be used for calibration multidecade IVDs including not only each individual decade´s settings but also combinations of decade settings in succession. The uncertainty of this method is estimated to be within 2×10-8 at 1 kHz and 2×10-7 at 10 kHz. An example of complete self-calibration for first and second decades of a multidecade IVD by the GIV method, and the corresponding data are given.
  • Keywords
    calibration; measurement standards; voltage dividers; 1 kHz; 10 kHz; decade settings; generalized injected voltage; measurement standards; multidecade inductive voltage divider; self-calibration; Calibration; Data processing; Electric potential; Metrology; Silicon; Standards; Uncertainty;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.1987.6312621
  • Filename
    6312621