DocumentCode
1296715
Title
A single-load microcalorimetric technique with newly designed X-band load for precise measurement of microwave power
Author
Agrawal, V.K. ; Kothari, P.C. ; Aggarwal, Ritander ; Bhatnagar, H.M.
Author_Institution
National Physical Laboratory, Hillside Road, New Delhi, India 110012
Issue
2
fYear
1987
fDate
6/1/1987 12:00:00 AM
Firstpage
450
Lastpage
454
Abstract
This paper describes a newly designed X-band calorimetrie load of high efficiency, greater than 99.9 percent, and the single-load microcalorimetric technique for precise measurement of microwave power. Because of the thin absorbing film used in the load, it is shown that the behavior of load with alternate application of dc and microwave power is almost identical. The results of microwave power measurements by the single-load microcalorimetric technique at 10.0 and 12.0 GHz, and their agreement with the power measured using thin-film barretter mounts, developed earlier, are given. The total uncertainty in power measurement is ±0.17 percent.
Keywords
Cooling; Electromagnetic heating; Gold; Microwave measurements; Microwave theory and techniques; Power measurement; Time measurement;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.1987.6312718
Filename
6312718
Link To Document