DocumentCode
1296752
Title
Calibration of a waveguide-below-cutoff attenuator using a voltage doubler
Author
Powell, Robert C.
Author_Institution
Weinschel Engineering Company, Inc., Gaithersburg, MD 20877; Powell Instruments, Gaithersburg, MD 20877
Issue
2
fYear
1987
fDate
6/1/1987 12:00:00 AM
Firstpage
480
Lastpage
486
Abstract
The design and operation of an automatic precision attenuator using a laser interferometer and a method for utilizing the new National Bureau of Standards (NBS) 1.25 MHz voltage doubler calibration system have been discussed previously [1], [2]. The purpose of this paper is to provide the technique required to determine corrections for multimode propagation and leakage in a waveguide below cutoff attenuation standard. Sufficient details are given to enable several orders of accuracy improvement near the attenuation limits. Calibrated values for skin-depth correction vary from prior publications [6]. Several areas requiring further research which were identified using these techniques are discussed.
Keywords
Approximation methods; Attenuation; Attenuation measurement; Attenuators; Calibration; Equations; NIST;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.1987.6312723
Filename
6312723
Link To Document