• DocumentCode
    1296813
  • Title

    On-line accuracy assessment for the dual six-port ANA: Extension to nonmating connectors

  • Author

    Hoer, Cletus A. ; Engen, Glenn F.

  • Author_Institution
    National Bureau of Standards, Boulder, CO 80303
  • Issue
    2
  • fYear
    1987
  • fDate
    6/1/1987 12:00:00 AM
  • Firstpage
    524
  • Lastpage
    529
  • Abstract
    A technique is described for calibrating a dual six-port or four-port automatic network analyzer (ANA) so that the scattering parameters of two-port devices having any combination of connectors can be measured. The technique is a generalization of the “thru-reflect-line” (TRL) calibration technique in which the thru is replaced with a second length of precision transmission line. Expressions for errors associated with the second line are derived.
  • Keywords
    Calibration; Connectors; Equations; Transmission line measurements; Uncertainty;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.1987.6312732
  • Filename
    6312732