  • Abstract
    The design principles of the Physikalisch-Technische Bundesanstalt´s (PTB´s) new primary clock, CS2, are reported. Experimental results concerning C field homogeneity, RF power dependence, cavity phase difference, and its gradient are given. The theoretical frequency instability due to shot and thermal noise, σ(2, 1 s) = 2.7 × 10−12, is compatible with the data so far obtained. The present uncertainty of CS2 is estimated to be 22 × 10−15. Frequency intercomparisons of the PTB´s two primary clocks, CSI, and CS2, averaged over one year yield a fractional difference of (3 ± 4) × 10−l5.
  • Keywords
    Cavity resonators; Clocks; Frequency measurement; Magnetic resonance; Magnetic shielding; Ovens; Uncertainty;