DocumentCode :
1302592
Title :
Authors´ reply
Author :
Hoffmann, Karel ; Skvor, Z.
Author_Institution :
Fac. of Electr. Eng., Czech Tech. Univ., Prague, Czech Republic
Volume :
58
Issue :
9
fYear :
2010
Firstpage :
2518
Lastpage :
2518
Abstract :
In order to obtain results in the above paper, we used five parameters. According to our statement that inserting a transmission line between the scalar analyzer and a perturbation two-port would not affect measured values, we imposed the condition that the argument of S11 is zero during calibration and data correction. Any other real value would do the same job.
Keywords :
network analysers; perturbation two-port; scalar analyzer; transmission line; vector network analyzer; Network analyzers;
fLanguage :
English
Journal_Title :
Microwave Theory and Techniques, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9480
Type :
jour
DOI :
10.1109/TMTT.2010.2058583