DocumentCode :
1304058
Title :
Correction of energy-dependent systematic errors in dual-energy X-ray CT using a basis material coefficients transformation method
Author :
Goh, K.L. ; Liew, S.C. ; Hasegawa, B.H.
Author_Institution :
Dept. of Phys., Nat. Univ. of Singapore, Singapore
Volume :
44
Issue :
6
fYear :
1997
fDate :
12/1/1997 12:00:00 AM
Firstpage :
2419
Lastpage :
2424
Abstract :
Computer simulation results from our previous studies showed that energy dependent systematic errors exist in the values of attenuation coefficient synthesized using the basis material decomposition technique with acrylic and aluminum as the basis materials, especially when a high atomic number element (e.g., iodine from radiographic contrast media) was present in the body. The errors were reduced when a basis set was chosen from materials mimicking those found in the phantom. In the present study, we employed a basis material coefficients transformation method to correct for the energy-dependent systematic errors. In this method, the basis material coefficients were first reconstructed using the conventional basis materials (acrylic and aluminum) as the calibration basis set. The coefficients were then numerically transformed to those for a more desirable set materials. The transformation was done at the energies of the low and high energy windows of the X-ray spectrum. With this correction method using acrylic and an iodine-water mixture as our desired basis set, computer simulation results showed that accuracy of better than 2% could be achieved even when iodine was present in the body at a concentration as high as 10% by mass. Simulation work had also been carried out on a more inhomogeneous 2D thorax phantom of the 3D MCAT phantom. The results of the accuracy of quantitation were presented here
Keywords :
calibration; computerised tomography; error correction; image reconstruction; measurement errors; medical image processing; 3D MCAT phantom; Al; I2-H2O; X-ray spectrum; accuracy; acrylic; aluminum; attenuation coefficient; basis material coefficients transformation method; basis material decomposition technique; calibration basis set; computer simulation results; dual energy X-ray CT; energy-dependent systematic errors; high atomic number element; high energy windows; inhomogeneous 2D thorax phantom; iodine; iodine-water mixture; low energy windows; phantom; quantitation; radiographic contrast media; Aluminum; Atomic measurements; Attenuation; Calibration; Computational modeling; Computer errors; Computer simulation; Error correction; Imaging phantoms; Radiography;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/23.656446
Filename :
656446
Link To Document :
بازگشت