• DocumentCode
    1304129
  • Title

    Study of multispectral frame-by-frame convolution scatter correction in high resolution PET

  • Author

    Yao, R. ; Bentourkia, M. ; Lecomte, R.

  • Author_Institution
    Dept. of Nucl. Med. & Radiobiol., Sherbrooke Univ., Que., Canada
  • Volume
    44
  • Issue
    6
  • fYear
    1997
  • fDate
    12/1/1997 12:00:00 AM
  • Firstpage
    2489
  • Lastpage
    2493
  • Abstract
    High resolution PET scanners based on small individual detectors have a low sensitivity resulting from increased inter-detector scatter. It has been postulated that this limitation can be overcome by using enlarged discrimination windows to include more low-energy events and by developing more efficient energy-dependent methods to correct for scatter. In this work, we investigate one such method based on the frame-by-frame scatter correction of multispectral data. Images acquired in the conventional (344-658 keV), broad (129-658 keV) and multispectral (129-658 keV) window modes were processed by stationary and nonstationary convolution-restoration scatter correction methods. Broad and multispectral window acquisition improved counting efficiency relative to the conventional window. The degradation of image quality due to added scatter can be mostly recovered by the subtraction-restoration scatter correction. However, the multispectral frame-by-frame method was found to be more sensitive to the effects of the nonstationary scatter response functions and its performance was not as good as that of the broad window. It is concluded that more sophisticated scatter degradation models and correction methods need to be established to take full advantage of multispectral data
  • Keywords
    convolution; gamma-ray detection; image resolution; image restoration; medical image processing; positron emission tomography; spectral analysis; 129 to 658 keV; 344 to 658 keV; PET scanners; broad window modes; conventional window modes; counting efficiency; energy-dependent methods; enlarged discrimination windows; high resolution PET; image quality degradation; inter-detector scatter; low sensitivity; low-energy events; multispectral frame-by-frame convolution scatter correction; multispectral window modes; nonstationary convolution-restoration scatter correction; stationary convolution-restoration scatter correction; subtraction-restoration scatter correction; Computational modeling; Convolution; Degradation; Detectors; Image quality; Kernel; Laboratories; Nuclear medicine; Positron emission tomography; Scattering;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.656457
  • Filename
    656457