DocumentCode :
1304648
Title :
Object Discrimination With an Artificial Hand Using Electrical Stimulation of Peripheral Tactile and Proprioceptive Pathways With Intrafascicular Electrodes
Author :
Horch, Kenneth ; Meek, Sanford ; Taylor, Tyson G. ; Hutchinson, Douglas T.
Author_Institution :
Dept. of Bioeng., Univ. of Utah, Salt Lake City, UT, USA
Volume :
19
Issue :
5
fYear :
2011
Firstpage :
483
Lastpage :
489
Abstract :
Trans-radial amputee subjects were implanted with intrafascicular electrodes in the stumps of the median and ulnar nerves. Electrical stimulation through these electrodes was used to provide sensations of touch and finger position referred to the amputated hand. Two subjects were asked to identify different objects as to size and stiffness by manipulating them with a myo-electric hand without visual or auditory cues. Both subjects were provided with information about contact force with the objects via tactile sensations referred to their phantom hands. One subject, who was provided with information about finger position in the prosthetic hand via a different tactile sensation referred to his phantom hand, was unable to correctly identify the objects. The other subject, who received information about finger position via a proprioceptive sensation referred to his phantom hand, correctly identified the objects at a level statistically significantly above chance performance.
Keywords :
artificial limbs; biomedical electrodes; neuromuscular stimulation; phantoms; touch (physiological); artificial hand; electrical stimulation; finger position; intrafascicular electrode; median nerve; myoelectric hand; object discrimination; object size; object stiffness; peripheral tactile pathway; phantom hand; proprioceptive pathway; proprioceptive sensation; prosthetic hand; tactile sensation; transradial amputee subjects; ulnar nerve; Cities and towns; Connectors; Electrodes; Fingers; Joints; Object recognition; Prosthetic hand; Amputee; intrafascicular electrodes; nerve stimulation; proprioception; prosthetic limb; touch; Adult; Amputation; Amputees; Artificial Limbs; Discrimination (Psychology); Electric Stimulation; Electrodes, Implanted; Electromyography; Feedback; Fingers; Hand; Humans; Male; Median Nerve; Neural Pathways; Peripheral Nerves; Phantom Limb; Proprioception; Touch; Ulnar Nerve;
fLanguage :
English
Journal_Title :
Neural Systems and Rehabilitation Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
1534-4320
Type :
jour
DOI :
10.1109/TNSRE.2011.2162635
Filename :
5995170
Link To Document :