DocumentCode :
1307089
Title :
Volitional Control of a Prosthetic Knee Using Surface Electromyography
Author :
Ha, Kevin H. ; Varol, Huseyin Atakan ; Goldfarb, Michael
Author_Institution :
Dept. of Mech. Eng., Vanderbilt Univ., Nashville, TN, USA
Volume :
58
Issue :
1
fYear :
2011
Firstpage :
144
Lastpage :
151
Abstract :
This paper presents a method for providing volitional control of a powered knee prosthesis during nonweight-bearing activity such as sitting. The method utilizes an impedance framework, such that the joint can be programmed with a given stiffness and damping that reflects the nominal impedance properties of an intact joint. Volitional movement of the knee joint is commanded via the stiffness set-point angle of the joint impedance, which is commanded by the user as a function of the measured surface electromyogram (EMG) from the hamstring and quadriceps muscles of the residual limb. Rather than using the respective EMG measurements from these muscles to directly command the flexion or extension set point of the knee, the presented approach utilizes a combination of quadratic discriminant analysis and principal component analysis to align the user´s intent to flex or extend the knee joint with the pattern of measured EMG. The approach was implemented on three transfemoral amputees, and their ability to control knee movement was characterized by a set of knee joint trajectory tracking tasks. Each amputee subject also performed the same set of trajectory tracking tasks with his sound side (intact) knee joint. The average root mean square trajectory tracking errors of the prosthetic knee employing the EMG-based volitional control and the intact knee of the three subjects were 6.2° and 5.2°, respectively.
Keywords :
artificial limbs; electromyography; medical control systems; medical signal processing; principal component analysis; damping; hamstring muscle; impedance framework; intact joint impedance properties; knee joint trajectory tracking tasks; nonweight bearing activity; powered knee prosthesis; principal component analysis; prosthetic knee volitional control; quadratic discriminant analysis; quadriceps muscle; residual limb; sEMG; sitting activity; stiffness set point angle; surface electromyography; transfemoral amputees; Electrodes; Electromyography; Impedance; Joints; Knee; Prosthetics; Trajectory; Electromyography; lower-limb prosthesis; myoelectric control; powered prosthesis; transfemoral prosthesis; Amputees; Electromyography; Humans; Knee; Knee Prosthesis; Quadriceps Muscle; Range of Motion, Articular; Signal Processing, Computer-Assisted;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.2010.2070840
Filename :
5559403
Link To Document :
بازگشت