• DocumentCode
    1307517
  • Title

    Novel Estimation of the Electrical Bioimpedance Using the Local Polynomial Method. Application to In Vivo Real-Time Myocardium Tissue Impedance Characterization During the Cardiac Cycle

  • Author

    Sanchez, Benjamin ; Schoukens, Johan ; Bragos, Ramon ; Vandersteen, Gerd

  • Author_Institution
    Dept. of Electron. Eng., Tech. Univ. of Catalonia, Barcelona, Spain
  • Volume
    58
  • Issue
    12
  • fYear
    2011
  • Firstpage
    3376
  • Lastpage
    3385
  • Abstract
    Classical measurements of myocardium tissue electrical impedance for characterizing the morphology of myocardium cells, as well as cell membranes integrity and intra/extra cellular spaces, are based on the frequency-sweep electrical impedance spectroscopy (EIS) technique. In contrast to the frequency-sweep EIS approach, measuring with broadband signals, i.e., multisine excitations, enables to collect, simultaneously, multiple myocardium tissue impedance data in a short measuring time. However, reducing the measuring time makes the measurements to be prone to the influence of the transients introduced by noise and the dynamic time-varying properties of tissue. This paper presents a novel approach for the impedance-frequency-response estimation based on the local polynomial method (LPM). The fast LPM version presented rejects the leakage error´s influence on the impedance frequency response when measuring electrical bioimpedance in a short time. The theory is supported by a set of validation measurements. Novel preliminary experimental results obtained from real-time in vivo healthy myocardium tissue impedance characterization within the cardiac cycle using multisine excitation are reported.
  • Keywords
    bioelectric phenomena; biological tissues; biomedical measurement; cardiology; cellular biophysics; electric impedance measurement; polynomial approximation; broadband signals; cardiac cycle; cell membrane integrity; electrical bioimpedance estimation; electrical impedance spectroscopy; extracellular spaces; fast LPM version; frequency sweep EIS; impedance frequency response estimation; in vivo real time myocardium electrical impedance; intracellular spaces; leakage error effects; local polynomial method; multiple myocardium tissue impedance data; multisine excitations; myocardium cell morphology; Frequency measurement; Frequency response; Impedance; Impedance measurement; Noise; Time measurement; Voltage measurement; Broadband electrical impedance spectroscopy (EIS); electrical bioimpedance (EBI); local polynomial method (LPM); multisine excitation; myocardium tissue electrical impedance; Animals; Dielectric Spectroscopy; Female; Heart; Models, Cardiovascular; Regression Analysis; Reproducibility of Results; Signal Processing, Computer-Assisted; Signal-To-Noise Ratio; Swine;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.2011.2166116
  • Filename
    5999711