• DocumentCode
    1310175
  • Title

    A new video-synchronized multichannel biomedical data acquisition system

  • Author

    Zeng, Shengke ; Powers, John R. ; Hsiao, Hongwei

  • Author_Institution
    Nat. Inst. for Occupational Safety & Health, Morgantown, WV, USA
  • Volume
    47
  • Issue
    3
  • fYear
    2000
  • fDate
    3/1/2000 12:00:00 AM
  • Firstpage
    412
  • Lastpage
    419
  • Abstract
    This data acquisition system records video frames onto a video tape, and simultaneously acquires biomedical data along with video time codes onto a computer hard disk to achieve a 30-min video-synchronized data recording with a summed data rate of 2.16 Mbit/s. A time-code-bridge file created during acquisition matches each video frame-start with the corresponding index number of the acquired data. The mean synchronization accuracy of the system is 0.22 ms.
  • Keywords
    biomedical equipment; data acquisition; medical computing; television applications; video recording; 0.22 ms; 30 min; 30-min video-synchronized data recording; acquisition matches; computer hard disk; mean synchronization accuracy; medical instrumentation; summed data rate; time-code-bridge file; video time codes; video-synchronized multichannel biomedical data acquisition system; Bandwidth; Bioinformatics; Biomedical computing; Data acquisition; Disk recording; Frequency synchronization; Hard disks; Occupational safety; Video recording; Videoconference; Computer Graphics; Computers; Data Collection; Environmental Monitoring; Humans; Information Storage and Retrieval; Medical Records Systems, Computerized; Monitoring, Physiologic; Signal Processing, Computer-Assisted; Software Design; Video Recording;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/10.827316
  • Filename
    827316