Bioimpedance spectroscopy tensor probe for anisotropic measurements
Author :
Kwon, Hyuck ; Wi, Hun ; Karki, B. ; Lee, E.J. ; McEwan, Alistair ; Woo, Eung Je ; Harrach, Bastian ; Seo, Jin Keun ; Oh, Tong
Author_Institution :
Dept. of Comput. Sci. & Eng., Yonsei Univ., Seoul, South Korea
Volume :
48
Issue :
20
fYear :
2012
Firstpage :
1253
Lastpage :
1255
Abstract :
The tetrapolar electrode configuration used in bioimpedance suffers from negative sensitivity and is confounded by anisotropic tissue such as blood vessels or muscle and nerve fibres. Proposed is a circular array of electrodes to focus current directly underneath the probe and provide anisotropy information. It is shown that an implementation using 16 miniature electrodes is able to outperform the tetrapolar method in sensitivity and impedance estimation. The proposed method can also recover anisotropic conductivity tensors from experiments for the first time.
Keywords :
bioelectric potentials; biomedical electrodes; blood vessels; electric impedance; electric impedance measurement; microelectrodes; muscle; neurophysiology; sensitivity; anisotropic conductivity tensors; anisotropic measurements; anisotropic tissue; anisotropy information; bioimpedance spectroscopy tensor probe; blood vessels; circular electrode array; impedance estimation; miniature electrodes; muscle; negative sensitivity; nerve fibres; tetrapolar electrode configuration;