DocumentCode :
1318700
Title :
Design for a constant peak current defibrillator
Author :
Machin, John Willis ; Brownhill, Julie ; Furness, Anthony
Author_Institution :
Dept. of Electr. & Electron. Eng., Staffordshire Polytech., Stafford, UK
Volume :
37
Issue :
7
fYear :
1990
fDate :
7/1/1990 12:00:00 AM
Firstpage :
723
Lastpage :
730
Abstract :
A damped sine wave defibrillator that is capable of delivering a constant peak discharge current over a wide range of patient thoracic impedance is shown to be theoretically possible. This is realized in practice by the technique of storing an amount of energy above that required for delivery, and dumping the excess charge when the discharge current has reached a predetermined level, the dumping being triggered by a current sensor. A larger than normal inductance assists in swamping the effect of varying impedance and produces a smooth waveform. The preselected peak current can be delivered with a variation of about +or-6% to 95% of a normal patient population. The energy expended in a simulated patient circuit is somewhat lower than for conventional defibrillators.
Keywords :
defibrillators; cardiac therapy; constant peak current defibrillator; current sensor; damped sine wave defibrillator; discharge current; patient thoracic impedance; simulated patient circuit; smooth waveform; Capacitors; Circuits; Defibrillation; Diodes; Heart; Immune system; Impedance measurement; Inductance; Shape control; Voltage; Computer Simulation; Electric Conductivity; Electric Countershock; Equipment Design; Humans; Male; Models, Biological; Thorax;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/10.55683
Filename :
55683
Link To Document :
بازگشت